Search
Browse Categories

Digital Pressure & Temperature Indicator/Alarm

Digital Pressure/Temperature Indicator, optional alarm outputs. Includes recorder output.

Price: $675.00
Item Number: GT439

Product Options

* denotes required field

Alarms*
Pressure Range*
Units*
Thermocouple*
Shipment in 2

Shipment in 2-3 Weeks Standard
Expedited delivery may be available

  • UNIVERSAL PRESSURE INDICATOR

    The ISE/Gentan model GT439 Digital Pressure/Temperature indicator/power supply is designed for use with any four leg, 350 ohm Wheatstone bridge pressure transducer. Zero and span can be easily adjusted without removing the module. Shunt calibration is a standard feature.

  • EASY TO READ LED DISPLAY

    The large 4½ digit, 1/3 inch LED pressure display is easy to read, even from a distance. Custom face plates are not required for specific scales because every unit can be set for 12 different ranges.  Other markings can be easily added for engineering units. The last digit is a fixed zero.

    A three digit LED temperature display is also included.

  • OPTIONAL DUAL ALARM SETPOINTS

    Two alarms can be selected for above or below setpoint control. Relays provide N.O. (normally open) and N.C. (normally closed) contacts for maximum flexibility. Adjustments are made through pots on the faceplate.

  • VERSATILE

    Provides a choice of recorder outputs, damping constant, transducer sensitivities, input power, shunt-resistor calibrations, peak-hold, and reset options.

  • INDUSTRY COMPATIBLE

Can be set to operate with Gentran or other styles of transducers.

  • EASY INSTALLATION

    The easy-to-mount aluminum housing fits a 1/4 DIN cutout. Slide-type metal brackets hold the case firmly in place and can be installed quickly. The electronic module can be easily removed for option selection or service

 Standard Pressure Ranges:

0-1,500 psig
0-3,000 psig
0-5,000 psig
0-10,000 psig
0-15,000 psig
 

 
 Accuracy:

Display: ±0.1% of full scale

Recorder Output (Pressure Only):  ±0.1% of full scale

Repeatability: ±0.1% of full scale
Linearity: ±0.1% of full scale
Stability: ±0.1% of full scale

 Physical Characteristics:

Size: 1/4 DIN
Display: 4½ digit LED, 1/3 inch height, last digit fixed zero and selectable decimal point
Operating Temperature: 0-140°F
Weight: 2.5 lbs.
 

 Power:

106-125 or 200-250 VAC, 50/60 hz, 1/8 Amp maximum
 

 Calibration:

Zero Adjustment: ±35% with front panel pot
Sensitivity Adjustment (Span): 1-2.5 mV/V or 2.5-5.0 mV/V, Switch selectable and fully adjustable.
Shunt Resistance Calibration: Selectable for Gentran or Dynisco
Response Time (Damping): Selectable at 0.3 or 5.0 seconds
 

 Recorder Output:

Standard: Selectable 0-1 VDC, 0-2 VDC, 0-5 VDC & 0-10 VDC
Optional: 4-20 mA, Load from 15-600 ohms

 Transducer Power Supply:
:8.2 VDC ±5.0%
 Optional Alarm Outputs:

Dual Relays: 2 SPDT (Single Pole Double Throw)
Relay Rating: 8 Amp @ 125/250 VAC, 5 Amp @ 30 VDC
Set Point Range: 1-100% of full scale
Accuracy: ±0.3% of full scale
Hysteresis: ±0.5% of full scale
Indication: Front panel LED's
Mode: Above or below set point
Reset: Automatic or manual
 

 Temperature Characteristics:

Thermocouple Types: J or K
Range: 0-999F or C
Accuracy: ±0.5%
 

.

 

Additional Information

Click Here to Download Product Literature
Shopping Cart
Your cart is empty.
Transit Times Via UPS